Japan JEOL JSM-7500F
1. Accelerating voltage: 0.3 - 30 kV
2. Lamp silk: W(310) cold field emission
3. Resolving power: 1.0 nm
4. Magnification: 25 – 1,000,000
5. Image capture: CCD-Camera
6. EDS (OXFORD INCA X-ACT)
7. Deben Centaurus Backscatter Detector.
Charging method
JEOL JSM-7500F
Fee Items CGU Labs / Non CGU Labs
Unused EDS 800 NT (hour) / 1600 NT (hour)
Use EDS 1000 NT (hour) / 2000 NT (hour)
Specimen Preparation Requirements:
7. Deben Centaurus Backscatter Detector.
Charging method
JEOL JSM-7500F
Fee Items CGU Labs / Non CGU Labs
Unused EDS 800 NT (hour) / 1600 NT (hour)
Use EDS 1000 NT (hour) / 2000 NT (hour)
Specimen Preparation Requirements:
- Please place the sample on an aluminum stub with a diameter of 26 mm.
- To maintain vacuum integrity, the sample must not decompose, release gas, or become liquid under electron beam exposure.
- To prevent damage, this instrument will not accept powder samples or materials that are strongly magnetic, magnetic (e.g., iron, cobalt, nickel), or easily attracted by the electromagnetic lens.
- To avoid damage, this instrument will not accept samples that are toxic, corrosive, volatile, or have a low melting point.
- To prevent computer contamination, please use a CD for burning electronic files.